AGENT SKILL

htsmlm

Operate htSMLM and EMU configurations through their semantic hardware mappings.

License 许可
BSD-3-Clause
最近更新
2026-10-20
Stars
★ 10
Forks
1

01 前言 Front Matter

name
htsmlm
description
Operate htSMLM and EMU configurations through their semantic hardware mappings.

03 SKILL.md 正文

htSMLM / EMU reference

Returned by load_skill(name="htsmlm").

What EMU and htSMLM are

EMU (Easier Micro-Manager User interfaces) is a Micro-Manager 2 plugin that provides configurable graphical user interfaces. The configuration wizard maps named "UI properties" (abstract controls in the GUI) to real Micro-Manager device properties on the running hardware. This mapping is saved as a JSON file at <MM_app_dir>/EMU/config.uicfg.

htSMLM is an EMU plugin designed specifically for SMLM microscopes. It provides:

  • Per-laser on/off and power controls (up to 4 lasers)
  • Filter wheel position controls
  • Focus panel with Z-stage positioning and focus-lock toggle
  • Quadrant photodiode (QPD) signal monitor
  • Laser trigger mode / pulse duration / sequence control (MicroFPGA)
  • Additional two-state device toggles (e.g., TIRF/HILO/3D/BFP modes)
  • Optional iBeamSmart laser panels
  • An Acquisition Wizard for unsupervised multi-position experiments

htSMLM/EMU is NOT always present. Only ask about or use it if the user mentions htSMLM, EMU, or a SMLM control GUI running inside Micro-Manager.


How to control htSMLM hardware via microclaw

Because EMU has no scripting API, control goes through the underlying Micro-Manager device properties that EMU is configured to use.

Workflow:

  1. Call get_emu_configuration() to read the UIProperty→MM device/property mapping from the EMU config file.
  2. Look up the UIProperty of interest in the returned "properties" dict. Each entry has: "device" — the MM device label (use with set_device_property) "property" — the MM property name (use with set_device_property) "on" — value for ON state (TwoState properties only) "off" — value for OFF state (TwoState properties only) "slope" — linear scale factor (Rescaled properties only) "offset" — linear offset (Rescaled properties only)
  3. Call set_device_property(device, property, value) with the device and property from step 2.

Configured human-facing identities ride on the semantic map records as name. Call resolve_emu_device with either an exact UIProperty key or a configured name such as "640" or "BFP". A laser name returns the whole slot record so its enable, power, and trigger lines remain paired.

Use get_emu_laser_map and resolve_emu_device instead of trial-and-error property probing. The semantic map already states which property is writable, the filter-wheel state table, and the focus-lock property.

Never calculate an EMU percentage conversion in prose. Call verify_emu_laser_power_calibration, then use set_emu_laser_power_percentage and get_emu_laser_power_percentage. If the requested, effective, measured, or GUI state disagree, or the requested value is not representable, keep illumination disabled and report the disagreement.

Example — enable Laser 0: config = get_emu_configuration() entry = config["properties"]["Laser 0 enable"] set_device_property(entry["device"], entry["property"], entry["on"])

Example — set Laser 0 to 50% power: verify_emu_laser_power_calibration(laser_slot=0, ...) set_emu_laser_power_percentage(laser_slot=0, percent=50) get_emu_laser_power_percentage(laser_slot=0)


UIProperty name inventory

These are the exact UIProperty names used in htSMLM. They appear as keys in the "properties" dict returned by get_emu_configuration().

Lasers (one panel per laser, labeled "Laser 0" through "Laser 3")

"Laser 0 enable" TwoState — laser on/off "Laser 0 power percentage" Rescaled — 0–100 power % (MM value scaled by slope/offset) (same pattern for Laser 1, 2, 3)

Filters (panel label "Filters")

"Filter wheel position" MultiState — filter wheel slot (0-based index) "Filter wheel 2 position" MultiState — second filter wheel (DualFW mode only)

Focus (panel label "Focus")

"Z stage position" SingleState — numeric Z position in µm "Z stage focus locking" TwoState — engage/disengage hardware focus lock

Additional two-state controls (panel label "Controls")

"Two-state device 1" through "Two-state device 6" These are mapped by the user to any two-state MM device property. Common uses: - TIRF vs. HILO vs. epi illumination mode - 3D astigmatic lens in/out - BFP (back focal plane / Bertrand lens) in/out - Focus lock laser on/off Their configured names are returned in each record's name field.

Laser triggers (panel label "Laser trigger 0" through "Laser trigger 3")

"Laser trigger 0 mode" MultiState — Off | On | Rising | Falling | Camera "Laser trigger 0 pulse duration" Rescaled — pulse length in µs (max 65535) "Laser trigger 0 sequence" MultiState — 16-bit binary pattern string (same pattern for triggers 1, 2, 3)

iBeamSmart panels (panel label is user-configurable, e.g. "405" or "iBeamSmart #1")

"<name> laser power" Rescaled — laser power "<name> operation" TwoState — on/off "<name> enable fine" TwoState — fine power mode on/off (optional) "<name> fine a (%)" Rescaled — fine A power (optional) "<name> fine b (%)" Rescaled — fine B power (optional)

QPD (panel label "QPD")

"QPD X", "QPD Y", "QPD Z" — read-only signal monitors; do not set these.

Powermeter (panel label "Powermeter")

"Laser powermeter" — read-only power readout; do not set this.


Plugin-level settings (from plugin_settings in get_emu_configuration)

These control which panels are visible. Useful for knowing what's available: "Powermeter tab" bool — powermeter tab enabled "Trigger tab" bool — laser trigger tab enabled (MicroFPGA) "Additional FW tab" bool — additional filter wheel tab enabled "Single FW panel" bool — true = single FW, false = dual FW "QPD tab" bool — QPD tab enabled "iBeamSmart #1" bool — first iBeamSmart panel enabled "iBeamSmart #1 name" str — display name of first iBeamSmart panel "iBeamSmart #2" bool — second iBeamSmart panel enabled "iBeamSmart #2 name" str — display name of second iBeamSmart panel "Additional FW tab title" str — title of the additional filters tab


Acquisition compatibility (examined 2026-09-08)

At htSMLM commit 30b6bfd (30b6bfd923a13076c08011ee001cc3a0e83c5434), LocalizationAcquisition starts ActivationTask and MMStudio MDA. ActivationTask analyses queued image pairs concurrently and publishes to an EMU property; there is no before-every-frame verification barrier. Its stop path polls at one-second intervals and waits a delay in seconds after its criterion. This implements neither a three-frame blink window nor exactly 1,000 frames after the terminal value. Do not delegate those requirements to this implementation. A later version can supersede these facts only after inspecting its analysis window and stop implementation; record the installed version and evidence.

For matching algorithms, check reachable instance, supported start/status/stop API, enforceable property/dose bounds, restoration, storage and standalone export before delegation. The shipped plugin hooks expose scalar analysis and autofocus, not a general activation controller; list_mm_plugins does not establish that the Acquisition Wizard is controllable. Use the acquisition tools for contracts those integrations cannot express.


Key questions to ask if the user has htSMLM

  1. What is the µManager installation directory? (Needed to read the EMU config; try get_emu_configuration() first — it will prompt for the directory if auto-detection fails.)

  2. Are the iBeamSmart panels enabled and what are their names? (Determines the UIProperty name prefix for those lasers.)

  3. Is the Trigger tab (MicroFPGA) present? (Required to set laser trigger modes and pulse durations.)