03 SKILL.md 正文
htSMLM / EMU reference
Returned by load_skill(name="htsmlm").
What EMU and htSMLM are
EMU (Easier Micro-Manager User interfaces) is a Micro-Manager 2 plugin that provides configurable graphical user interfaces. The configuration wizard maps named "UI properties" (abstract controls in the GUI) to real Micro-Manager device properties on the running hardware. This mapping is saved as a JSON file at <MM_app_dir>/EMU/config.uicfg.
htSMLM is an EMU plugin designed specifically for SMLM microscopes. It provides:
- Per-laser on/off and power controls (up to 4 lasers)
- Filter wheel position controls
- Focus panel with Z-stage positioning and focus-lock toggle
- Quadrant photodiode (QPD) signal monitor
- Laser trigger mode / pulse duration / sequence control (MicroFPGA)
- Additional two-state device toggles (e.g., TIRF/HILO/3D/BFP modes)
- Optional iBeamSmart laser panels
- An Acquisition Wizard for unsupervised multi-position experiments
htSMLM/EMU is NOT always present. Only ask about or use it if the user mentions htSMLM, EMU, or a SMLM control GUI running inside Micro-Manager.
How to control htSMLM hardware via microclaw
Because EMU has no scripting API, control goes through the underlying Micro-Manager device properties that EMU is configured to use.
Workflow:
- Call get_emu_configuration() to read the UIProperty→MM device/property mapping from the EMU config file.
- Look up the UIProperty of interest in the returned "properties" dict. Each entry has: "device" — the MM device label (use with set_device_property) "property" — the MM property name (use with set_device_property) "on" — value for ON state (TwoState properties only) "off" — value for OFF state (TwoState properties only) "slope" — linear scale factor (Rescaled properties only) "offset" — linear offset (Rescaled properties only)
- Call set_device_property(device, property, value) with the device and property from step 2.
Configured human-facing identities ride on the semantic map records as name.
Call resolve_emu_device with either an exact UIProperty key or a configured
name such as "640" or "BFP". A laser name returns the whole slot record so its
enable, power, and trigger lines remain paired.
Use get_emu_laser_map and resolve_emu_device instead of trial-and-error property
probing. The semantic map already states which property is writable, the
filter-wheel state table, and the focus-lock property.
Never calculate an EMU percentage conversion in prose. Call
verify_emu_laser_power_calibration, then use
set_emu_laser_power_percentage and get_emu_laser_power_percentage. If the
requested, effective, measured, or GUI state disagree, or the requested value
is not representable, keep illumination disabled and report the disagreement.
Example — enable Laser 0: config = get_emu_configuration() entry = config["properties"]["Laser 0 enable"] set_device_property(entry["device"], entry["property"], entry["on"])
Example — set Laser 0 to 50% power: verify_emu_laser_power_calibration(laser_slot=0, ...) set_emu_laser_power_percentage(laser_slot=0, percent=50) get_emu_laser_power_percentage(laser_slot=0)
UIProperty name inventory
These are the exact UIProperty names used in htSMLM. They appear as keys in the "properties" dict returned by get_emu_configuration().
Lasers (one panel per laser, labeled "Laser 0" through "Laser 3")
"Laser 0 enable" TwoState — laser on/off "Laser 0 power percentage" Rescaled — 0–100 power % (MM value scaled by slope/offset) (same pattern for Laser 1, 2, 3)
Filters (panel label "Filters")
"Filter wheel position" MultiState — filter wheel slot (0-based index) "Filter wheel 2 position" MultiState — second filter wheel (DualFW mode only)
Focus (panel label "Focus")
"Z stage position" SingleState — numeric Z position in µm "Z stage focus locking" TwoState — engage/disengage hardware focus lock
Additional two-state controls (panel label "Controls")
"Two-state device 1" through "Two-state device 6"
These are mapped by the user to any two-state MM device property. Common uses:
- TIRF vs. HILO vs. epi illumination mode
- 3D astigmatic lens in/out
- BFP (back focal plane / Bertrand lens) in/out
- Focus lock laser on/off
Their configured names are returned in each record's name field.
Laser triggers (panel label "Laser trigger 0" through "Laser trigger 3")
"Laser trigger 0 mode" MultiState — Off | On | Rising | Falling | Camera "Laser trigger 0 pulse duration" Rescaled — pulse length in µs (max 65535) "Laser trigger 0 sequence" MultiState — 16-bit binary pattern string (same pattern for triggers 1, 2, 3)
iBeamSmart panels (panel label is user-configurable, e.g. "405" or "iBeamSmart #1")
"<name> laser power" Rescaled — laser power "<name> operation" TwoState — on/off "<name> enable fine" TwoState — fine power mode on/off (optional) "<name> fine a (%)" Rescaled — fine A power (optional) "<name> fine b (%)" Rescaled — fine B power (optional)
QPD (panel label "QPD")
"QPD X", "QPD Y", "QPD Z" — read-only signal monitors; do not set these.
Powermeter (panel label "Powermeter")
"Laser powermeter" — read-only power readout; do not set this.
Plugin-level settings (from plugin_settings in get_emu_configuration)
These control which panels are visible. Useful for knowing what's available: "Powermeter tab" bool — powermeter tab enabled "Trigger tab" bool — laser trigger tab enabled (MicroFPGA) "Additional FW tab" bool — additional filter wheel tab enabled "Single FW panel" bool — true = single FW, false = dual FW "QPD tab" bool — QPD tab enabled "iBeamSmart #1" bool — first iBeamSmart panel enabled "iBeamSmart #1 name" str — display name of first iBeamSmart panel "iBeamSmart #2" bool — second iBeamSmart panel enabled "iBeamSmart #2 name" str — display name of second iBeamSmart panel "Additional FW tab title" str — title of the additional filters tab
Acquisition compatibility (examined 2026-09-08)
At htSMLM commit 30b6bfd (30b6bfd923a13076c08011ee001cc3a0e83c5434),
LocalizationAcquisition starts ActivationTask and MMStudio MDA. ActivationTask
analyses queued image pairs concurrently and publishes to an EMU property;
there is no before-every-frame verification barrier. Its stop path polls at
one-second intervals and waits a delay in seconds after its criterion. This
implements neither a three-frame blink window nor exactly 1,000 frames after
the terminal value. Do not delegate those requirements to this implementation.
A later version can supersede these facts only after inspecting its analysis
window and stop implementation; record the installed version and evidence.
For matching algorithms, check reachable instance, supported start/status/stop API, enforceable property/dose bounds, restoration, storage and standalone export before delegation. The shipped plugin hooks expose scalar analysis and autofocus, not a general activation controller; list_mm_plugins does not establish that the Acquisition Wizard is controllable. Use the acquisition tools for contracts those integrations cannot express.
Key questions to ask if the user has htSMLM
-
What is the µManager installation directory? (Needed to read the EMU config; try get_emu_configuration() first — it will prompt for the directory if auto-detection fails.)
-
Are the iBeamSmart panels enabled and what are their names? (Determines the UIProperty name prefix for those lasers.)
-